| Insulation
Resistance |
| Test
Voltage |
250V/500V/1000V |
Measuring
Ranges
(Mid-scale Value) |
| 0~100MΩ/ |
0~200MΩ/ |
0~400MΩ |
| (1MΩ) |
(2MΩ) |
(4MΩ) |
|
Output
Voltage
on open circuit |
Rated
test voltage ±10% |
| Rated
Current |
0.8mA
DC approx. |
Output
Short
Circuit Current |
2mA
DC approx. |
Accuracy
Guaranteed
Ranges |
0.05~10MΩ/0.1~20MΩ/0.2~40MΩ |
| Accuracy |
|±5% of indicated value |
| Continuity
Test |
Measuring
Ranges
(Mid-scale Value) |
50Ω
(20Ω) |
Output
Voltage
on open circuit |
0.6V
DC approx. |
Output
Short
Circuit Current |
300mA
approx. |
| Accuracy |
|±3% of scale length |
| General |
| Withstand
Voltage |
2200V AC for
1 minute |
| Power
Source |
R6P(AA)(1.5V)
× 8 |
| Dimensions |
143(L) ×
93(W) × 63(D)mm |
| Weight |
500g approx. |
| Accessories |
7025(Test
leads)
fuse 250V/1A
9038(Carrying Case)
R6P(AA) × 8
Instruction Manual |